Thermo Fisher Genetic Analyzers Allow DNA File Tampering

Published Vorant ratingroutine vulnerability Healthcare

Vorant's rating weighs the evidence in the report: confirmed exploitation, victims claimed, CVSS and EPSS, and how many sources agree. It rates the report, not your assets.

A missing integrity check in Thermo Fisher Applied Biosystems genetic analyzer software lets attackers modify .fsa/.hid files, corrupting DNA test results.

Sourcecisa.gov — open the original

What to check

Applies to
The 1 CVE named below. Whether you run an affected version is not assessed: check the versions the source lists.
Exploitation
No exploitation in the wild and no public exploit is recorded for the CVEs named.

CISA published an advisory for multiple Thermo Fisher Applied Biosystems genetic analyzer products, disclosing a vulnerability (CVE-2026-17583) tied to CWE-353, Missing Support for Integrity Check. The affected software fails to protect .fsa/.hid output files from unauthorized modification, meaning an attacker with access to these files could alter DNA data and produce inaccurate test outcomes. The affected product line spans multiple Data Collection Software versions used in the Applied Biosystems 3500/3500xL, 3730/3730xL, SeqStudio, SeqStudio Flex, GeneMapper ID-X, 3130 Series, and legacy ABI PRISM instruments, deployed worldwide in healthcare and public health laboratories.

Thermo Fisher has released patched versions for several actively supported products that add digital signatures to instrument software, allowing verification that data files have not been tampered with. However, three older product lines (3130 Series, ABI PRISM 3100/3100-Avant, and ABI PRISM 310) are end-of-life and will not receive fixes, leaving those deployments reliant on interim mitigations such as chain-of-custody controls, encrypted storage, least-privilege access, and network isolation.

CISA notes the vulnerability is not remotely exploitable and no public exploitation has been reported. The impact is primarily one of data integrity in forensic and clinical DNA analysis workflows rather than a network intrusion vector, but given the sensitivity of DNA testing (forensic, paternity, clinical diagnostics), tampering could have significant downstream consequences if exploited by an insider or someone with local file access.

Mentioned in this report

Vulnerabilities CVE-2026-17583

Source reporting: https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-216-01

Free public brief. Slack alerts on the vendors you run come with the paid plans, and a new account gets three days of them, no card.

What this brief leaves out

This page is the free tier: the write-up, the severity, the names. In the app the same report carries live profiles of the CVEs it names. The paid plans add the security research and news reporting the free tier leaves out.

It sits in a corpus of 16,445 reports from 149 sources, 666 of them written in the last seven days, and it grows through the day.

A new account starts with three days of all of it, no card, then keeps the government and CERT reporting free. Slack alerts are a paid-plan feature: name the vendors you run and Slack hears about the reports that touch them, not the rest of the week's. Nothing is charged unless you subscribe.

Create a free account   What it costs